Hach China
400-686-8899 / 800-840-6026
 

CA610 and storage

產(chǎn)品分類:CA610在線氟化物分析儀
更新日期:2018-11-09
● 問題咨詢

CA610 and storage

● 關(guān)鍵詞摘要

CA610,在線氟化物

● 解答

To store the CA610, all the lines should be flushed with DI water and electrodes removed. Then either empty the fluoride electrode and store dry or place the electrodes in a chlorinated sample to keep them conditioned for when it is time to put them back online. If they are stored dry, the electrodes will need to be reconditioned. If stored wet, just keep the lower 1 inch of the electrodes immersed in water. Loosen the pinch block to keep the tubing from being crushed.

這個解答對您是否有幫助?


評價

 

您的瀏覽器版本過低,存在重大安全隱患,請升級到IE10以上,或使用最新版Chrome、Firefox瀏覽,謝謝