SP510 Maintenance Requirements
SP510,水質(zhì)硬度檢測儀
Colorimeter needs to be cleaned monthly or as needed
Reagents
Replace buffer and indicator reagents every two months.
Discard old containers and any unused contents.
Standardization
Standardization must be performed after reagent installation or when an optical system component has been replaced.
Replace pump tubing
Every 6 months < 27oC
Every 3 months > 27oC
Replace remaining instrument tubing annually
Hach has a complete portfolio of instruments and chemistries with support and services to help you get the right results.