Hach China
400-686-8899 / 800-840-6026
 

SS7 Verification

產(chǎn)品分類:Surface Scatter 7sc 高量程濁度儀
更新日期:2018-11-12
● 問題咨詢

SS7 Verification

● 關(guān)鍵詞摘要

SS7,濁度,高量程濁度儀

● 解答

Perform a calibration
Confirm to perform a baseline using a standardization plate.
The serial number for the last standardization plate used will appear on the display. Confirm to accept or enter the serial number of the plate to be used and confirm.
Follow the controller prompts:
Remove the calibration cylinder and wipe off the top of the sample cylinder.
Place the standardization plate on top of the sample cylinder so that the light beam strikes the center of the plate. Note the orientation of the plate and always place it in the same position when using it to check standardization.
Close the door to eliminate stray light. Confirm to continue.
When the measured value becomes stable, confirm to establish an expected value for the plate.
Note: Future measured values will be compared to the stored expected value. If the established PASS criteria are not met, a new calibration should be performed.

這個(gè)解答對(duì)您是否有幫助?


評(píng)價(jià)

 

您的瀏覽器版本過低,存在重大安全隱患,請(qǐng)升級(jí)到IE10以上,或使用最新版Chrome、Firefox瀏覽,謝謝